QBC STAR™ POC Haematology System

QBC BIG.JPG 

The QBC STAR™ POC Haematology System is the simplest solution for patient side testing. An innovative approach employing dry haematology reagents instead of the bulky reagents utilised by other methods. A revolution in patient care- easier, more efficient and virtually problem-free.

Measures the Clinically Significant Parameters Most Requested by Physicians:

  • haematocrit
  • haemoglobin
  • MCHC
  • platelet count
  • white blood cell count
  • granulocyte percentage
  • lymphocyte/monocyte count
  • lymphocyte/monocyte percentage

FEATURES.JPG

Re-read Functionality:

The QBC STAR™ now comes with re-read functionality allowing the user to re-read a sample that has already been spun and to access precision as well as accuracy. Troubleshooting is alos made easier as a tube can be removed from the STAR™, assessed and then placed in the analyser again for a second run

Patient Identification System:

Patient indentification information can be entered simply via the keyboard and barcode scanner

Click here for full download

KEY.JPG            
  

 Operating Ranges

Haematology Parameters measured with the QBC STAR™ system are valid over the following range of values:

  • Haematocrit          0.15 - 0.65
  • Haemoglobin          5.0 - 20.0 g/dL
  • Platelet Count        20 - 999 x 10⁹/L
  • WBC Count            1.6 - 99.9 x 10⁹/L
  • Granulocyte Count  0.8 - 70.0 x 10⁹/L
  • Lymph/Mono Count 0.8 - 99.9 x 10⁹/L

 Accuracy, Precision and Comparative Data

Table 1 - Accuracy comparing the QBC STAR™ system with the Coulter® STKS or Sysmex™ K1000.
ParameterCorrelation CoefficientSlopeInterceptQBC MeanCell Counter MeanRange of ValuesNumber of Samples
Haematocrit (%)0.9830.9732.57236.534.815.7-61.7        646
Haemoglobin (g/dL)0.9840.9820.38712.112.05.2-18.5638
platelet (x 10⁹/L)  0.9620.93517.70124424223-913558
WBC (x 10⁹/L)0.9741.124-.93610.410.11.6-92.9535
Gramulocyte  (x 10⁹/L)0.9720.9910.1527.07.00.08-45.0535
Lymph/Mono (x10⁹/L)0.9871.206-.4193.33.10.8-89.9535

 

Table 2 - Accuracy comparing the QBC STAR™ system against the international microhaematocrit reference method
ParameterCorreclation CoefficientSlopeInterceptQBC MeanCell Counter MeanRange of ValuesNumber of Samples
Microhaematocrit (%)0.9861.023-.65036.536.315.7-61.9646

 

Table 3 - Precision data on typical within-run precision tests in the QBC STAR™ system are shown in the two tables below. The precision data represent the analysis of eleven whole blood specimens, each assayed in replicates of ten                                                                                                                
ParameterMean ValueMean % CV
HCT (%)41.72.0%
HB (g/dL)14.01.9%
PLT (x10⁹/L)2356.0%
WBC (x10⁹/L)6.06.4%
ParameterRangeMx S.D
GRAN (%)38-793.2
LYMPHO/MONO (%)21-633.2

 