The QBC STAR™ POC Haematology System is the simplest solution for patient side testing. An innovative approach employing dry haematology reagents instead of the bulky reagents utilised by other methods. A revolution in patient care- easier, more efficient and virtually problem-free. Measures the Clinically Significant Parameters Most Requested by Physicians: - haematocrit
- haemoglobin
- MCHC
- platelet count
- white blood cell count
- granulocyte percentage
- lymphocyte/monocyte count
- lymphocyte/monocyte percentage
| 
|
Re-read Functionality: The QBC STAR™ now comes with re-read functionality allowing the user to re-read a sample that has already been spun and to access precision as well as accuracy. Troubleshooting is alos made easier as a tube can be removed from the STAR™, assessed and then placed in the analyser again for a second run Patient Identification System: Patient indentification information can be entered simply via the keyboard and barcode scanner Click here for full download | |
| | |
Operating Ranges
Haematology Parameters measured with the QBC STAR™ system are valid over the following range of values:
- Haematocrit 0.15 - 0.65
- Haemoglobin 5.0 - 20.0 g/dL
- Platelet Count 20 - 999 x 10⁹/L
- WBC Count 1.6 - 99.9 x 10⁹/L
- Granulocyte Count 0.8 - 70.0 x 10⁹/L
- Lymph/Mono Count 0.8 - 99.9 x 10⁹/L
Accuracy, Precision and Comparative Data
Table 1 - Accuracy comparing the QBC STAR™ system with the Coulter® STKS or Sysmex™ K1000.| Parameter | Correlation Coefficient | Slope | Intercept | QBC Mean | Cell Counter Mean | Range of Values | Number of Samples |
| Haematocrit (%) | 0.983 | 0.973 | 2.572 | 36.5 | 34.8 | 15.7-61.7 | 646 |
| Haemoglobin (g/dL) | 0.984 | 0.982 | 0.387 | 12.1 | 12.0 | 5.2-18.5 | 638 |
| platelet (x 10⁹/L) | 0.962 | 0.935 | 17.701 | 244 | 242 | 23-913 | 558 |
| WBC (x 10⁹/L) | 0.974 | 1.124 | -.936 | 10.4 | 10.1 | 1.6-92.9 | 535 |
| Gramulocyte (x 10⁹/L) | 0.972 | 0.991 | 0.152 | 7.0 | 7.0 | 0.08-45.0 | 535 |
| Lymph/Mono (x10⁹/L) | 0.987 | 1.206 | -.419 | 3.3 | 3.1 | 0.8-89.9 | 535 |
Table 2 - Accuracy comparing the QBC STAR™ system against the international microhaematocrit reference method| Parameter | Correclation Coefficient | Slope | Intercept | QBC Mean | Cell Counter Mean | Range of Values | Number of Samples |
| Microhaematocrit (%) | 0.986 | 1.023 | -.650 | 36.5 | 36.3 | 15.7-61.9 | 646 |
Table 3 - Precision data on typical within-run precision tests in the QBC STAR™ system are shown in the two tables below. The precision data represent the analysis of eleven whole blood specimens, each assayed in replicates of ten | Parameter | Mean Value | Mean % CV |
| HCT (%) | 41.7 | 2.0% |
| HB (g/dL) | 14.0 | 1.9% |
| PLT (x10⁹/L) | 235 | 6.0% |
| WBC (x10⁹/L) | 6.0 | 6.4% |
| Parameter | Range | Mx S.D |
| GRAN (%) | 38-79 | 3.2 |
| LYMPHO/MONO (%) | 21-63 | 3.2 |